A neonate may need to be transferred from one place to another for providing medical care. Any major disturbance during transport can be harmful to the baby. HB2000 is an incubator transporting babies safely and without disturbance.
1. Microprocessor based servo control mode: air temperature, skin temperature.
2. Alternating AC power supply and DC power supply, DC12V or DC 24V power supply is used on the ambulance.
3.LED display: air temperature, skin temperature, set temperature and internal battery level.
4. Self-test function; Various audible and visual fault alerts.
5. The crib can be easily pulled out of the side door.
6. Incubator height is adjustable.
7. Natural airflow humidity and oxygen supply system.
8. Optional: >37 degree, overrid temperature mode
Standard configuration:
Main body(including the hood, controller, bassinet, internal battery, and observation lamp) oxygen cylinders, oxygen supply system, skin temperature sensor, IV pole, mattress, adjustable stand.
AC power supply: | AC220V-230V/50HZ |
DC power supply: | DC12V/10A or DA24V/6A |
Input power: | 400VA |
Model of control: | air mode and baby model controlled by micro-computer |
Air temperature control range:
override mode: |
25℃-37℃
37℃-38℃ |
Baby temperature control range:
override mode: |
34℃-37℃
37℃-37.5℃ |
Skin temperature sensor accuracy: | ±0.2℃ |
Temperature variability: | ≤1.0 |
Temperature uniformity: | ≤1.5℃(mattress is on the horizontal position) |
Noise inside hood: | ≤52dB(A) |
Failure alarm: | over-temperature-alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on |
Operating at temperature equilibrium: | 90min(set temp 36℃ ambient 15℃) |
Environmental temperature?operating range: | 10℃-30℃(set temp should be higher 3℃ than environmental temp) |
Ambient air movement rate: | 〉0.3m/s,〈1.0m/s |