• Hospital Defibrillator, AED Defibrillator, Medical Defibrillator, Portable Defibrillator, Automated Defibrillator, Medical Defibrillator, Defibrillator Use In Hospital
  • Hospital Defibrillator, AED Defibrillator, Medical Defibrillator, Portable Defibrillator, Automated Defibrillator, Medical Defibrillator, Defibrillator Use In Hospital

Defi5 Hospital Portable AED Automated External Defibrillator

An automated external defibrillator (AED) is a lightweight, portable device that delivers an electric shock through the chest to the heart. The shock can potentially stop an irregular heart beat (arrhythmia) and allow a normal rhythm to resume following sudden cardiac arrest (SCA).

  • Product Description
  • Technical Parameters
  • Video

Automatic External Defibrillator Features include:
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
Physical

Dimensions 303 x 216 x 89 mm
Weight  2.0 kg
Operating Temperature 0℃ to 40℃
Operating Humidity Relative humidity between 30% and 95% (non-condensing)
Storage Temperature (without battery)  -20℃ to 55℃
Storage Humidity(without battery) Up to 93% (non-condensing)

Electrical Isolation

Power Unit operates on internal battery only
External Electrical Connections No external devices are attached to the unit
Risk Current Category Internally powered equipment with defibrillator-proof BF type

Defibrillator

Biphasic Truncated exponential Less than 30 seconds
150, 150, 200J Less than 35 seconds
10 sec. to 150J Less than 25 seconds
15 sec. to 200J 22 audible prompts
9 sec. LED prompts

Waveform details
The table below provides details of the biphasic truncated exponential waveform delivered by the DEFI5 (set to 200J) when connected to resistive loads of 25 through 175 Ohms. The waveforms are characterized by typical values for peak current (Ip), duration of the first output phase, and duration of the second output phase. Values are within 10%.
Battery

Non-rechargeable 12V d.c. 2.8Ah
Capacity 100 discharges at 200 Joules or 120 discharges at 150 Joules
Shelf Life (25℃±15℃) 5 years

Hospital Defibrillator, AED Defibrillator, Medical Defibrillator, Portable Defibrillator, Automated Defibrillator, Medical Defibrillator, Defibrillator Use In Hospital

Output

Energy(J)

Patient

Impedance(Ω)

I p1

(Amps)

I p2

(Amps)

Phase1

(ms)

Phase2

(ms)

Interval

(ms)

200 25 53.6 36.0 5.5 3.3 0.7
50 27.8 17.5 10.9 6.6 0.7
75 18.2 12.0 16.3 8.9 0.7
100 13.4 9.0 19.7 9.6 0.7
125 10.5 7.2 20.5 11.2 0.7
150 8.6 6.0 21.6 10.4 0.7
175 8.2 6.1 21.8 11.2 0.7

Video