Zhangjiagang City, Jiangsu Province, China.
+86-512-58810663
An electrocardiogram records the electrical signal from your heart to check for different heart conditions. Electrodes are placed on your chest to record your heart’s electrical signals, which cause your heart to beat. The signals are shown as waves on an attached computer monitor or printer.
ECG1202 Portable ECG Monitor Features:
Simultaneous 12 leads ECG data acquisition and display
Simultaneous 12 leads ECG printing with 210mm/216mm thermal paper (Roll or Z-fold)
10 inch 800 x 480 TFT color LCD
Alphanumeric keyboard
Built-in rechargeable lithium battery, print up to 360 ECG reports (optional)
Built-in 4GB SD card, save up to 10,000 ECG reports in XML format (optional)
Pacemaker detection
Precise ECG auto-interpretation USB SLAVE port can be connected to PC-ECG software (optional)
Connect to an external USB printer supporting PCL6 language by USB MASTER port to print12 leads ECG report on A4 paper
Network transmission (optional)
Software upgrade automatically via SD card
Lead | Standard 12 leads |
Input Circuit | Built-in defibrillator protection |
Acquisition Mode | Simultaneous12 leads acquisition |
Measurement Range | ±5mVpp |
Time Constant | ≥3.2s |
Frequency Response | 0.05Hz~150Hz (-3dB) |
Sensitivity | 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV |
Input Impedance | ≥50MΩ |
Input Circuit Current | <0.05µA |
Noise Level | <15µVp-p |
Anti Baseline Drift | Automatic |
Patient Current Leakage | <10µA(220V-240V) |
Filter | EMG Filter:25 Hz/ 30 Hz/ 40Hz |
Low Pass Filter:75 Hz/ 100 Hz/ 150Hz | |
DFT Filter:0.05 Hz/ 0.15 Hz/ 0.25 Hz/ 0.32 Hz/ 0.5 Hz/ 0.67Hz/ ADS | |
AC Filter : 50 Hz / 60Hz | |
CMRR | 98dB |
Automatic: 12CH×1, 6CH×2, 6CH×2+1R, 3CH×4, 3CH×4+1R ,3CH×4+3R | |
Recording Mode | Manual: 12CH, 6CH, 3CH, 2CH, 1CH |
Rhythm: Any lead selectable | |
Recording System | Thermal matrix printing |
Recording Paper | Roll: 216mm×30m, 210mm×30m |
Z-fold: 210mm×295mm×200 sheets | |
Paper Speed | 5 mm/s, 6.25 mm/s, 10 mm/s, 25 mm/s, 50 mm/s |
Display | 800×480 pixel TFT LCD |
Interface | Network interface, USB Master/Slave, SD card |
Safety Standard | IEC I/CF |
Sample Rate | Normal: 1000sps/channel |
Pacemaker: 10,000sps/channel | |
Pacemaker Detection | ≥± 2 mV ≥± 0.1 ms |
Dimension | 435mm×335mm×98mm |
Weight | 4Kg(include battery) |
AC: 100~240V, 50/60Hz, 70VA~100VA | |
Power Supply | DC: 14.8V/4400mAh, built-in lithium battery |